Validity and intra-rater reliability of an Android phone application to measure cervical range-of-motion

机译:用于测量子宫颈活动范围的Android手机应用程序的有效性和评估者内部可靠性

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

Background: Concurrent validity and intra-rater reliability using a customized Android phone application to measure cervical-spine range-of-motion (ROM) has not been previously validated against a gold-standard three-dimensional motion analysis (3DMA) system. Findings. Twenty-one healthy individuals (age:31 ± 9.1 years, male:11) participated, with 16 re-examined for intra-rater reliability 1-7 days later. An Android phone was fixed on a helmet, which was then securely fastened on the participantu27s head. Cervical-spine ROM in flexion, extension, lateral flexion and rotation were performed in sitting with concurrent measurements obtained from both a 3DMA system and the phone.The phone demonstrated moderate to excellent (ICC = 0.53-0.98, Spearman ρ = 0.52-0.98) concurrent validity for ROM measurements in cervical flexion, extension, lateral-flexion and rotation. However, cervical rotation demonstrated both proportional and fixed bias. Excellent intra-rater reliability was demonstrated for cervical flexion, extension and lateral flexion (ICC = 0.82-0.90), but poor for right- and left-rotation (ICC = 0.05-0.33) using the phone. Possible reasons for the outcome are that flexion, extension and lateral-flexion measurements are detected by gravity-dependent accelerometers while rotation measurements are detected by the magnetometer which can be adversely affected by surrounding magnetic fields. Conclusion: The results of this study demonstrate that the tested Android phone application is valid and reliable to measure ROM of the cervical-spine in flexion, extension and lateral-flexion but not in rotation likely due to magnetic interference. The clinical implication of this study is that therapists should be mindful of the plane of measurement when using the Android phone to measure ROM of the cervical-spine.
机译:背景:使用定制的Android手机应用程序来测量颈椎活动范围(ROM)的并发有效性和评估者内部可靠性以前尚未针对黄金标准的三维运动分析(3DMA)系统进行过验证。发现。 21名健康人(年龄:31±9.1岁,男性:11岁)参加了试验,其中1至7天后对16名受试者的信度进行了重新检查。将Android手机固定在头盔上,然后将其牢固地固定在参与者的头上。坐着进行颈椎ROM的屈曲,伸展,横向屈曲和旋转,同时从3DMA系统和手机获得测量结果,手机表现出中等至出色的表现(ICC = 0.53-0.98,Spearmanρ= 0.52-0.98) ROM测量在颈屈,伸直,侧屈和旋转中的同时有效性。但是,颈椎旋转既显示比例偏差又显示固定偏差。事实证明,使用该手机对颈椎屈伸,侧屈和侧屈(ICC = 0.82-0.90)具有出色的评估者内部可靠性,但对于左右旋转(ICC = 0.05-0.33)则较差。结果的可能原因是屈曲,伸展和横向屈曲测量值由重力相关的加速度计检测,而旋转测量值由磁力计检测,这可能会受到周围磁场的不利影响。结论:这项研究的结果表明,经过测试的Android手机应用程序可以有效,可靠地测量颈椎的ROM的屈曲,伸展和横向屈曲,但由于磁性干扰而不能旋转。这项研究的临床意义在于,治疗师在使用Android手机测量颈椎ROM时应注意测量平面。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号