Multiple ion detection system with miniscan facilities and expanded mass range for magnetic sector mass spectrometers.

机译:具有miniscan设施和扩展的质量范围的多离子检测系统,适用于磁性扇形质谱仪。

摘要

A new four-channel multiple ion detection (MID) system is described. By adding a A V of up to 1100 volts to the nominal accelerating voltage ( V ) , a mass range of about 3 0 % ( V = 3600 volts) to 9 0% ( V = 1200 volts) can be covered, thus providing a remarkably wide range compared to other MID units. The system incorporates a ramp voltage, which added to V allows a small scan around each one of the peaks fo- udcused, as well as holding amplifiers in each channel. These features combined with the excellent stability of the Ionization udchamber against voltage changes mlnlmlre the errors due to defocusing effects, permitting the addition of voltages as high as those described. The scan time is continuously adjustable between 0.5 and 12 s and the swltching time between two adjacent channels is 50 ms. The circuit design, its operation, udand some aspects of the performance of this unit are described.
机译:描述了一种新的四通道多离子检测(MID)系统。通过将最高1100伏的AV添加到额定加速电压(V),可以覆盖大约3 0%(V = 3600伏)至9 0%(V = 1200伏)的质量范围,从而提供了显着的与其他MID单元相比,范围更广。该系统包含一个斜坡电压,该电压加到V上,可以对所聚焦的每个峰值进行较小的扫描,并在每个通道中保持放大器。这些特征与电离室对电压变化的出色稳定性相结合,从而消除了由于散焦效应引起的误差,从而允许增加如上所述的电压。扫描时间可在0.5至12 s之间连续调节,两个相邻通道之间的切换时间为50 ms。描述了电路设计,其工作原理和本机性能的某些方面。