Investigation of hemodynamics in an in vitro system simulating left ventricular support through the right subclavian artery using 4-dimensional flow magnetic resonance imaging.

机译:使用4维流磁共振成像模拟体外系统通过右锁骨下动脉对左心室支持的血液动力学研究。

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

OBJECTIVES\ud\udLeft ventricular assist devices are an important treatment option for patients with heart failure alter the hemodynamics in the heart and great vessels. Because in vivo magnetic resonance studies of patients with ventricular assist devices are not possible, in vitro models represent an important tool to investigate flow alterations caused by these systems. By using an in vitro magnetic resonance-compatible model that mimics physiologic conditions as close as possible, this work investigated the flow characteristics using 4-dimensional flow-sensitive magnetic resonance imaging of a left ventricular assist device with outflow via the right subclavian artery as commonly used in cardiothoracic surgery in the recent past.\ud\udMETHODS\ud\udAn in vitro model was developed consisting of an aorta with its supra-aortic branches connected to a left ventricular assist device simulating the pulsatile flow of the native failing heart. A second left ventricular assist device supplied the aorta with continuous flow via the right subclavian artery. Four-dimensional flow-sensitive magnetic resonance imaging was performed for different flow rates of the left ventricular assist device simulating the native heart and the left ventricular assist device providing the continuous flow. Flow characteristics were qualitatively and quantitatively evaluated in the entire vessel system.\ud\udRESULTS\ud\udFlow characteristics inside the aorta and its upper branching vessels revealed that the right subclavian artery and the right carotid artery were solely supported by the continuous-flow left ventricular assist device for all flow rates. The flow rates in the brain-supplying arteries are only marginally affected by different operating conditions. The qualitative analysis revealed only minor effects on the flow characteristics, such as weakly pronounced vortex flow caused by the retrograde flow via the brachiocephalic artery.\ud\udCONCLUSIONS\ud\udThe results indicate that, despite the massive alterations in natural hemodynamics due to the retrograde flow via the right subclavian and brachiocephalic arteries, there are no drastic consequences on the flow in the brain-feeding arteries and the flow characteristics in the ascending and descending aortas. It may be beneficial to adjust the operating condition of the left ventricular assist device to the residual function of the failing heart.
机译:目的\ ud \ ud左心室辅助设备是心力衰竭患者改变心脏和大血管的血流动力学的重要治疗选择。由于无法在室内进行带有心室辅助设备的患者的磁共振研究,因此体外模型代表了研究由这些系统引起的血流变化的重要工具。通过使用尽可能模拟生理状况的体外磁共振兼容模型,这项工作使用左心室辅助装置的4维流敏磁共振成像技术研究了流动特性,该设备通常通过右锁骨下动脉流出一种体外模型是由主动脉组成,其主动脉上分支与左心室辅助装置相连,该装置模拟自然衰竭心脏的搏动流。第二个左心室辅助装置通过右锁骨下动脉为主动脉提供连续的血流。针对模拟天然心脏的左心室辅助设备和提供连续血流的左心室辅助设备的不同流速执行了四维流敏磁共振成像。在整个血管系统中对流动特性进行定性和定量评估。\ ud \ udRESULTS \ ud \ ud主动脉及其上部分支血管内部的流动特性表明右锁骨下动脉和右颈动脉仅由左连续流支撑各种流量的心室辅助装置。大脑供血动脉中的流速仅受不同操作条件的影响很小。定性分析显示,对流动特性的影响很小,例如通过头臂动脉逆行流动引起的弱明显涡流。\ ud \ ud结论\ ud \ ud结果表明,尽管自然血液动力学因通过右锁骨下和头臂动脉的逆行血流,不会对脑喂食动脉的血流以及升主动脉和降主动脉的血流特征产生严重影响。将左心室辅助装置的操作条件调整为衰竭心脏的残余功能可能是有益的。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号