The influence of sample freezing at -80°C for 2-12 weeks on glycated haemoglobin (HbA1c) concentration assayed by HPLC method on Bio-Rad D-10© auto-analyzer.

机译:在Bio-Rad D-10©自动分析仪上通过HPLC法测定的-80°C冷冻2-12周的样品对糖化血红蛋白(HbA1c)浓度的影响。

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

Introduction: The aim of the study was to evaluate the effect of a single freeze/thaw cycle on HbA1c concentrations measured by commercially available HPLC method.Materials and methods: Study included 128 whole blood samples collected from diabetic patients (N = 60) and healthy volunteers (N = 68). HbA1c concentrations were measured in fresh blood samples. Then samples were frozen at - 80 °C for up to 12 weeks. HbA1c was assayed by ion-exchange HPLC method on Bio-Rad D-10® analyzer. Variables were compared using Wilcoxon and ANOVA Kruskal-Wallis tests. Bias between HbA1c measured in fresh and frozen samples was calculated. The comparability of HbA1c concentrations was assessed by Bland-Altman plot.Results: Median (IQR) HbA1c concentration was 45.3 (36.6–61.2) mmol/mol for fresh and 45.3 (36.6–60.6) mmol/mol for frozen/thawed samples. No significant difference in HbA1c concentrations was found comparing fresh and frozen/thawed samples (P = 0.070) in the whole group, as well as in healthy and diabetic subjects. The median calculated bias between fresh and frozen/thawed samples was 0% in whole group and healthy subjects, and 1.19% in diabetic patients. No significant difference was found between the biases according to baseline HbA1c values (P = 0.150). The Bland-Altman plot analysis showed a positive bias of 0.4% (95% CI: - 2.8 - 3.7%), which indicates high compliance between HbA1c values and no relevant influence of sample freezing on clinical significance of HbA1c measurement.Conclusions: Storage for up to 12 weeks at – 80 °C with a single freeze/thaw cycle does not affect HbA1c concentrations measured with HPLC method on Bio-Rad D-10® analyzer.
机译:简介:本研究的目的是评估单次冷冻/解冻循环对通过市售HPLC方法测量的HbA1c浓度的影响。材料和方法:研究包括128例从糖尿病患者(N = 60)和健康人那里采集的全血样品志愿者(N = 68)。在新鲜血液样本中测量HbA1c浓度。然后将样品在-80°C下冷冻12周。在Bio-RadD-10®分析仪上通过离子交换HPLC法测定HbA1c。使用Wilcoxon和ANOVA Kruskal-Wallis检验比较变量。计算在新鲜和冷冻样品中测得的HbA1c之间的偏差。结果:HbA1c浓度的中位数(IQR)对新鲜的HbA1c浓度为45.3(36.6–61.2)mmol / mol,对于冷冻/解冻的样品为45.3(36.6–60.6)mmol / mol。在整个组以及健康和糖尿病受试者中,比较新鲜和冷冻/融化的样品(P = 0.070),发现HbA1c浓度无显着差异。在整个组和健康受试者中,新鲜和冷冻/解冻样品之间的中值计算偏差为0%,而糖尿病患者为1.19%。根据基线HbA1c值,在偏倚之间未发现显着差异(P = 0.150)。 Bland-Altman图分析显示正偏差为0.4%(95%CI:-2.8-3.7%),这表明HbA1c值之间的一致性很高,并且样品冻结对HbA1c测量的临床意义没有相关影响。在单个冷冻/融化循环中,在– 80°C下长达12周不会影响通过Bio-RadD-10®分析仪上的HPLC方法测量的HbA1c浓度。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号