首页> 外文OA文献 >An evaluation of pocket-model, numerical readout breath alcohol testing instruments
【2h】

An evaluation of pocket-model, numerical readout breath alcohol testing instruments

机译:袖珍型数字读数呼吸酒精测试仪的评估

摘要

Eight small-scale breath alcohol measurement devices were tested for accuracy, precision and the ability to not yield false positive and false negative readings. These pocket-sized breath testers (PMBTs), which provided numerical readout of BrAC to the 100th of a percent, were smaller than evidential and preliminary breath test instruments (EBTs and PBTs). The smallest devices were approximately the same size of a cigarette lighter. Designed to provide drinkers feedback about their individual alcohol levels, the PMBTs ranged in price from $40-100 USD. The devices were first tested under laboratory conditions with alcohol solution simulators providing the alcoholic samples. They were then tested with human drinkers, under controlled field conditions. Each device was tested at multiple alcohol levels. Two of the eight PMBTs failed to complete all levels of testing and were excluded from the study. All PMBTs demonstrated the ability to not yield false positive and false negative readings. No device met NHTSA performance criteria for accuracy (systematic error) in testing EBTs at every alcohol level tested. An interaction between PMBTs and the alcohol test levels was found. Thus, accuracy was found to be dependent upon the alcohol level at which the devices were tested. No device met NHTSA performance criteria for precision in testing EBTs at every alcohol level tested. Precision varied depending on the testing condition. There was less precision under controlled field conditions than under laboratory conditions. Five of the six PMBTs that completed the testing overestimated BrAC; only one device read below actual BrAC. Ramifications of the findings are discussed, regarding the overestimation and underestimation of BrAC and the possibility of manufacturers intentionally calibrating the devices to overestimate BrAC. Potential PMBT users are discussed and areas for future research are addressed.
机译:测试了八种小型呼吸酒精测量设备的准确性,精密度以及不产生假阳性和假阴性读数的能力。这些袖珍型呼气测试仪(PMBT)可以将BrAC的数值读出到百分之一的100,比证据性和初步的呼气测试仪(EBT和PBT)要小。最小的设备大约与打火机的大小相同。 PMBT旨在向饮酒者提供有关其各自酒精含量的反馈,其价格范围从40美元至100美元不等。首先在实验室条件下用提供酒精样品的酒精溶液模拟器测试设备。然后在受控的野外条件下,对人类饮酒者进行了测试。每个设备都在多种酒精水平下进行了测试。八个PMBT中有两个未能完成所有级别的测试,因此被排除在研究之外。所有PMBT均具有不产生假阳性和假阴性读数的能力。在每种酒精含量下测试EBT的准确性(系统误差)都没有设备符合NHTSA性能标准。发现PMBT与酒精测试水平之间存在相互作用。因此,发现准确性取决于测试装置的酒精水平。没有一种设备能够满足NHTSA的性能标准,从而无法精确测试每种酒精含量的EBT。精度取决于测试条件。在受控的野外条件下,精度要比实验室条件下的精度低。完成测试的六个PMBT中有五个被高估了。仅一台设备的读数低于实际BrAC。讨论了发现的结果,涉及BrAC的高估和低估以及制造商有意校准设备以高估BrAC的可能性。讨论了潜在的PMBT用户,并讨论了未来的研究领域。

著录项

  • 作者

    Van Tassel William Edward;

  • 作者单位
  • 年度 2004
  • 总页数
  • 原文格式 PDF
  • 正文语种 en_US
  • 中图分类

相似文献

  • 外文文献
  • 中文文献
  • 专利

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号