Improved visualization of breast tissue on a dedicated breast PET system through ergonomic redesign of the imaging table

机译:通过符合成像台的符合人体工程学重新设计,改善了专用乳房宠物系统上的乳房组织的可视化

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

Abstract Background Breast lesions closer than 2 cm to the chest wall are difficult to position in the field of view of dedicated breast PET (db-PET) systems. This inability to detect such lesions is a significant limitation of these systems. The primary objective of this study was to determine if modifications to the design of the imaging table and detector used for a db-PET system would enable improved visualization of breast tissue close to the chest wall. All studies were performed on a commercially available db-PET system (Mammi-PET). A central square section of the imaging table, containing the standard 180-mm circular aperture, was modified such that it could be removed and replaced by thinner sections with a larger aperture. Additional changes were made to the cover plate of the detector array and the patient mattress. A total of 60 patients were studied. After administration of F-18 FDG, 30 patients were imaged with a 220-mm-diameter aperture and the standard aperture, and 30 patients with a 200-mm aperture and the standard aperture. On all scans, the length of breast tissue in the field of view was measured as the greatest extent of tissue from the nipple back to the posterior edge of the breast. Image quality and patient comfort were recorded. Results Averaged over both breasts, relative to the standard aperture, the increase in breast length was 12.5 + 7.7 mm with the 220-mm aperture, and 12.3 + 6.5 mm with the 200-mm aperture (p < 0.05 for both apertures). In ~ 5% of cases, the larger apertures resulted in some degradation in image quality due to closer proximity to cardiac/hepatic activity. In 10–20% of cases, movement of the breast tissue was observed as the detector ring was moved to scan the anterior region of the breast. The patient survey indicated no significant difference in the comfort level between the standard aperture and either of the prototype apertures. Conclusions Modifications to the image table and system resulted in a significant gain in the volume of breast tissue that could be imaged on the db-PET system and should allow better visualization of lesions close to the chest wall.
机译:抽象背景乳腺癌病灶距离小于2cm至胸壁难以在视专用乳房PET(DB-PET)系统的领域位置。这种无法检测这种病变是这些系统的显著限制。本研究的主要目的是,以确定是否修改用于分贝-PET系统的成像台和检测器的设计将实现改进的乳房组织接近胸壁的可视化。所有研究均市售的DB-PET系统(Mammi-PET)上进行。中央方形截面成像台的,含有标准的180毫米圆形孔径,被修改,使得它可以被除去,并通过与一个较大的光圈较薄部分取代。附加变化​​到检测器阵列的在盖板和患者床垫制造。总共有60名患者进行了研究。 F-18 FDG的给药后,30名患者用一个220毫米直径的孔,并且所述标准孔径,而30名患者用200毫米孔径和孔径的标准成像。上所有的扫描,乳房组织中的视场的长度测量为组织的从乳头回乳房的后边缘的最大程度。图像质量和患者的舒适度进行记录。结果进行平均超过两个乳房,相对于标准孔径,在乳房长度的增加是12.5 +7.7毫米与220毫米孔径,和12.3 +6.5毫米与200毫米孔(P <0.05为两个孔)。在箱子〜5%时,较大的孔导致的图像质量劣化的一些由于更接近心脏/肝活性。在箱子10-20%,观察到乳房组织的运动作为检测器环被移动到扫描乳房的前部区域。患者的调查显示在标准孔之间和任一的原型孔的舒适程度无差异显著。结论修改到图像表和系统导致乳房组织的体积可以在DB-PET系统上成像并应该允许病变更好的可视化接近胸壁一个显著增益。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号