摘要

A detection system for the continuous and simultaneous measurement of gas-phase chlorine and hydrogen chloride in the low parts per billion (109) by volume (ppbv) range is described. Both gases are trapped through the walls of a microporous polypropylene tube into an appropriate flowing recipient buffer. The buffer then flows through two electrochemical detectors placed in series downstream from the sampling tube. Chloride ions from gas-phase hydrogen chloride are detected potentiometrically with an Ag/AgCl working electrode. Chlorine is detected biamperometrically (at 0.1 V) via its oxidation of added iodide ions to form triiodide. In a continuous-flow measurement mode, detection of chloride and hydrogen chloride at levels as low as 0.75 and 2.1 ppbv, respectively, is possible. Use of a 2-min stopped-flow/flow-injection arrangement results in a three-fold improvement in detection limits. The selectivity of each detector with respect to carbon dioxide, sulfur dioxide, nitrogen dioxide and hydrogen sulfide is also examined.
机译:描述了用于在每十亿百亿(109)的低压分别(PPBV)范围内连续和同时测量气相氯和氯化氢的检测系统。将两种气体通过微孔聚丙烯管的壁捕获到合适的流动受体缓冲液中。然后缓冲器流过两个电化学探测器,从采样管中串联串联。用Ag / AgCl工作电极对来自气相氯化氢的氯离子进行滴度。通过其氧化加入的碘离子以形成三碘化碘,检测氯丁酸葡萄酸(0.1V)。在连续流动测量模式下,可以分别检测氯化物和氯化氢,分别为0.75和2.1ppbv的水平。使用2分钟的停止流动/流喷射装置导致检测限的三倍改善。还检查了各检测器相对于二氧化碳,二氧化硫,二氧化氮和硫化氢的选择性。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号