首页> 外文OA文献 >Leukocyte count and two-part differential in whole blood based on a portable microflow cytometer
【2h】

Leukocyte count and two-part differential in whole blood based on a portable microflow cytometer

机译:基于便携式微流式细胞仪的全血白细胞计数和两部分鉴别

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

This work demonstrated leukocyte count and two-part leukocyte differential from the whole blood based on a portable microflow cytometer system. Leukocytes are selectively stained with a fluorescent dye, Acridine Orange (AO). The blood sample is then pumped through a disposable microfluidic device for fluorescence sensing. Under blue LED excitation (460nm), the green fluorescence from DNA and the red fluorescence from RNA are detected simultaneously with two photomultiplier tubes (PMT). Leukocytes are counted and differentiated into two parts, lymphocyte versus non-lymphocyte, based on their fluorescence signatures. The results, including leukocyte absolute count and lymphocyte ratio, matched well with the commercial hemacytometer counts (maximal error 9.0%, correlation coefficient ~0.8). We also tested the system for the first time under a zero-gravity (zero-G) environment that facilitated its application in space missions.
机译:这项工作证明了基于便携式微流式细胞仪系统的白细胞计数和全血的两部分白细胞差异。白细胞用荧光染料A啶橙(AO)选择性染色。然后将血样泵送通过一次性微流体装置进行荧光传感。在蓝色LED激发(460nm)下,用两个光电倍增管(PMT)同时检测DNA的绿色荧光和RNA的红色荧光。对白细胞进行计数,并根据其荧光特征将其分为淋巴细胞与非淋巴细胞两部分。结果,包括白细胞绝对计数和淋巴细胞比率,与商业血细胞计数器计数非常吻合(最大误差9.0%,相关系数〜0.8)。我们还首次在零重力(zero-G)环境下测试了该系统,该系统有助于将其应用于太空任务。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号