摘要

We present an integrated photoacoustic and ultrasonic three-dimensional (3-D) volumetric imaging system based on a two-dimensional (2-D) matrix array ultrasound probe. A wavelength-tunable dye laser pumped by a Q-switched Nd:YAG laser serves as the light source and a modified commercial ultrasound imaging system (iU22, Philips Healthcare) with a 2-D array transducer (X7-2, Philips Healthcare) detects both the pulse-echo ultrasound and photoacoustic signals. A multichannel data acquisition system acquires the RF channel data. The imaging system enables rendering of co-registered 3-D ultrasound and photoacoustic images without mechanical scanning. The resolution along the azimuth, elevation, and axial direction are measured to be 0.69, 0.90 and 0.84 mm for photoacoustic imaging. In vivo 3-D photoacoustic mapping of the sentinel lymph node was demonstrated in a rat model using methylene blue dye. These results highlight the clinical potential of 3-D PA imaging for identification of sentinel lymph nodes for cancer staging in humans.
机译:我们提出了一个基于二维(2-D)矩阵阵列超声探头的集成的光声和超声三维(3-D)体积成像系统。由Q开关Nd:YAG激光器泵浦的波长可调染料激光器用作光源,带有二维阵列换能器(X7-2,Philips Healthcare)的改进型商用超声成像系统(iU22,Philips Healthcare)可以检测脉冲回波超声和光声信号都可以。多信道数据获取系统获取RF信道数据。该成像系统无需机械扫描即可渲染共同注册的3D超声和光声图像。对于光声成像,沿方位角,仰角和轴向的分辨率测量为0.69、0.90和0.84 mm。在使用亚甲基蓝染料的大鼠模型中,对前哨淋巴结进行了体内3-D光声映射。这些结果突出了3-D PA成像在识别人类癌症分期中前哨淋巴结方面的临床潜力。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号