The consequences of consuming and commercializing illicit drugs including cocaine, con­stitute a serious problem for authorities and the whole society. Cocaine is usually identified in the laboratory conditions by chromatographic or spectroscopic methods. Electro­che­mical techniques have also gained prominence because they are fast and easy to use, have many applications, and provide reproducible and reliable results. Therefore, in the present study, a voltammetric method was developed to detect freebase cocaine using carbon paste electrode and methanol as the main cocaine solvent. The developed method was applied to detect cocaine in the artificial saliva by the square wave voltammetry (SWV). The current values increased linearly with the concentration of cocaine, which afforded construction of the analytical curve. The limit of detection (LoD) and the limit of quantify­cation (LoQ) were determined as 0.90 µg/mL and 2.41 µg/mL, respectively. For compa­rison purposes, HPLC-DAD chromatographic method was also applied to detect cocaine. The corresponding analytical curve gave LoD = 0.043 µg/mL and LoQ = 0.130 µg/mL. Although showing better analytical results, HPLC-DAD method could not detect cocaine in saliva samples without previous treatment, what makes the electrochemical method much more attractive for this type of detection.
机译:消费和商业化包括可卡因的非法药物的后果构成了当局和整个社会的严重问题。可卡因通常通过色谱或光谱方法在实验室条件下鉴定。电化学技术也取得了突出,因为它们快速且易于使用,具有许多应用,并提供可重复和可靠的结果。因此,在本研究中,开发了伏安法,用于使用碳糊电极和甲醇作为主要可卡因溶剂来检测自由base可卡因。应用开发方法以通过方波伏安法(SWV)检测人造唾液中的可卡因。随着可卡因的浓度,电流值随着可卡因的浓度而增加,这得到了分析曲线的构建。检测极限(LOD)和量子次数(LOQ)的极限分别测定为0.90μg/ ml和2.41μg/ ml。为了比较目的,HPLC-DAD色谱法也应用于检测可卡因。相应的分析曲线给出LOD =0.043μg/ mL,LOQ =0.130μg/ mL。虽然显示出更好的分析结果,但HPLC-DAD方法在没有先前治疗的情况下无法检测到唾液样品中的可卡因,是什么使得电化学方法对这种类型的检测更具吸引力。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号