Accuracy of densitometry of two cone beam computed tomography equipment in comparison with computed tomography

机译:与计算机断层扫描相比,两个锥形光束计算机断层扫描设备的密度测定精度

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

Background: The use of oral implants has been growing, and cone beam computerized tomography (CBCT) hasbecome the method of choice for oral and maxillofacial radiology.Objective: To assess the accuracy of bone densitometry in two different CBCT devices in comparison withMDCT (multi-detector CT).Methods: Different concentrations of urografin, including 2.5%, 5%, 7.5%, 10%, 12.5%, were prepared, and theHounsfield unit of these solution was measured by two CBCT devices (SORDEX CRANEX 3D and NEWTOM5G) and one spiral CT device (SOMATOM SENSATION). Difference of output Hounsfield units in eachconcentration was compared in three devices. Correlation of devices with increase of urografin dose also wasevaluated. Statistical analyses of the data were performed using SPSS18 and Kruskal–Wallis and Mann–WhitneyU tests, along with Spearman’s correlation coefficient.Results: The range of gray density for NEWTOM 5G CBCT, SORDEX 3D CBCT, and SOMATOM CTimaging systems was from 781 to 2311, 427 to 1464, and 222 to 994, respectively. There was significantdifference between devices in the Hounsfield unit in all urografin concentrations (p<0.001). Also there was asignificant correlation between three devices with increasing the urografin dose (p0.95)Conclusion: Our findings indicated a high correlation and linear relationship between different studied imagingsystems. Although utilizing CBCT in the assessment of bone density is useful according to its lower emitted doseand less cost, clinicians should be aware of the issue that the voxel value in CBCT is not as perfect as CT.
机译:背景:口腔植入物的使用已经增长,锥梁电脑断层扫描(CBCT)具有成为口腔和颌面放射学的首选方法。目的:评估两种不同CBCT设备中骨密度测定的准确性与MDCT(多探测器CT)。方法:制备不同浓度的甲丙嘧啶,其中包括2.5%,5%,7.5%,10%,12.5%,和这些解决方案的Hounsfield单元由两个CBCT设备(Sordex Cranex 3D和Newtom)测量5G)和一个螺旋CT器件(体重感觉)。每个输出Hounsfield单位的差异在三个装置中比较浓度。装置随着丙比素剂量增加的相关性也是如此评估。使用SPSS18和Kruskal-Wallis和Mann-Whitney进行数据的统计分析U测试,以及Spearman的相关系数。结果:Newtom 5G CBCT,Sordex 3D CBCT和SOMATOM CT的灰度密度范围成像系统分别为781至2311,427至1464,以及222至994。有重要意义所有Urografin浓度的Hounsfield单元中的装置之间的差异(P <0.001)。还有一个三种器件随着Urografin剂量的显着相关性(P0.95)结论:我们的研究结果表明了不同研究成像之间的高相关和线性关系系统。虽然在评估骨密度中使用CBCT可根据其较低的发射剂量而有用并且成本较低,临床医生应该了解CBCT中的体素值并不像CT那样完美。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号