首页> 外文OA文献 >Study of the reference standard facility of rhodium-iron resistance thermometer
【2h】

Study of the reference standard facility of rhodium-iron resistance thermometer

机译:铑铁电阻温度计参考标准设备的研究

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

A 704 type ⁴He cryostat is designed and the “National (China) Temperature Working Standard Group of the ITS-90 at 1.2–24 K” have been composed. A 703 type ³He cryostat have been established and the working temperature of the “National Temperature Working Standard Group of the ITS-90 at 1.2–24 K” is extended to 0.65 K, which is the lowest limit temperature of ITS-90. The 5 rhodium-iron resistance thermometers (RIRTs) which compose the Working Standard Group have been transformed to the International Temperature Scale of 1990 (ITS-90). A set of indexing data is fitted and calculated as the indexing table of the standard facility of RIRT. 42 basis temperature points have been measured at 0.67–26.37 K. The results show that the temperature control level could reach 0.5 mK/30 min. The difference between the temperature value of the 5 RIRTs and the average is less than 1.1 mK, and the comparison measurements uncertainty is 0.95 mK.
机译:设计了704型⁴He低温恒温器,并组成了“ ITS-90的国家(中国)温度工作标准组(1.2–24 K)”。已经建立了703型“ He低温恒温器”,“ ITS-90的国家温度工作标准组在1.2–24 K下”的工作温度扩展到0.65 K,这是ITS-90的最低极限温度。组成工作标准组的5个铑铁电阻温度计(RIRT)已转换为1990年国际温度标度(ITS-90)。拟合并计算出一组索引数据,作为RIRT标准设施的索引表。测得的42个基本温度点为0.67–26.37K。结果表明,温度控制水平可以达到0.5 mK / 30分钟。 5个RIRT的温度值与平均值之间的差小于1.1 mK,比较测量不确定度为0.95 mK。

著录项

  • 作者

    Yin L.; Lin P.; Qi X.;

  • 作者单位
  • 年度 2014
  • 总页数
  • 原文格式 PDF
  • 正文语种 en
  • 中图分类

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号