首页> 外文OA文献 >Validation of handheld fundus camera with mydriasis for retinal imaging of diabetic retinopathy screening in China: a prospective comparison study
【2h】

Validation of handheld fundus camera with mydriasis for retinal imaging of diabetic retinopathy screening in China: a prospective comparison study

机译:验证手持式眼压相机与中国糖尿病视网膜病筛查视网膜病变的疗程:预期比较研究

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

Objectives To investigate the clinical validity of using a handheld fundus camera to detect diabetic retinopathy (DR) in China.Design and settings Prospective comparison study of the handheld fundus camera with a standard validated instrument in detection of DR in hospital and a community screening clinic in Guangdong Province, China.Participants Participants aged 18 years and over with diabetes who were able to provide informed consent and agreed to attend the dilated eye examination with handheld tests and a standard desktop camera.Primary and secondary outcome measures Primary outcome was the proportion of those with referable DR (R2 and above) identified by the handheld fundus camera (the index test) compared with the standard camera. Secondary outcome was the comparison of proportion of gradable images obtained from each test.Results In this study, we examined 304 people (608 eyes) with each of the two cameras under mydriasis. The handheld camera detected 119 eyes (19.5%) with some level of DR, 81 (13.3%) of them were referable, while the standard camera detected 132 eyes (21.7%) with some level of DR and 83 (13.7%) were referable. It seems that the standard camera found more eyes with referable DR, although McNemar’s test detected no significant difference between the two cameras.Of the 608 eyes with images obtained by desktop camera, 598 (98.4%) images were of sufficient quality for grading, 12 (1.9%) images were not gradable. By the handheld camera, 590 (97.0%) were gradable and 20 (3.2%) images were not gradable.The two cameras reached high agreement on diagnosis of retinopathy and maculopathy at all the levels of retinopathy.Conclusion Although it could not take the place of standard desktop camera on clinic fundus examination, the handheld fundus camera showed promising role on preliminary DR screening at primary level in China. To ensure quality images, mydriasis is required.
机译:目的探讨使用手持式眼底照相机来检测中国糖尿病视网膜病变(DR)的临床有效性。设计和设置手持眼底相机的预测比较研究标准验证仪器检测医院博士和社区筛查诊所中国广东省。糖尿病患者参与者18岁及以上的糖尿病,能够提供知情同意,并同意通过手持式测试和标准桌面相机参加扩张的眼科检查。提高和二次结果措施主要结果是那些比例与标准相机相比,通过手持式USFUS相机(索引测试)识别的可引用DR(R2及以上)。二次结果是从每次测试中获得的渐变图像比例的比较。在本研究中,我们将304人(608只眼睛)用两种摄像机进行了两种相机。手持式摄像机检测到119只眼睛(19.5%),其中一定程度的DR,81(13.3%)是可指选的,而标准摄像机检测到132只眼睛(21.7%),有一定程度的DR和83(13.7%)是可指选的。似乎标准相机发现更多的眼睛与可引用的DR,尽管麦克马尔的测试检测到两个相机之间没有显着差异。通过台式摄像头获得的图像,598(98.4%)图像具有足够的评分质量,12 (1.9%)图像不渐变。通过手持式摄像头,590(97.0%)渐变,20(3.2%)图像不渐变。两台相机达到高度协议,达到视网膜病变水平的视网膜病变和小疗病的诊断。结论虽然它无法取代作者:张莹莹,掌声,掌上眼底相机在中国初步研讨会上对初步博士筛查的初步研讨会展示了有前途的作用。为确保质量图像,需要迈索里索。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号