In the article, the specially designed devices for pre-sterilization treatment of laparoscopic, endoscopic and other long-length surgical equipment from organic compounds (blood and alkali) in ultrasonic tank-container with ozone has been proposed. The first device was equipped with a bath (up to 30 l), ten ultrasonic emitters with a total power 300 W, and four modular ozone generators based on dielectric barrier discharges. The second device equipped with 3.5 liter volume bath was also designed having a glass pipe 1 meter long, 8 cm in diameter with one ozone module and immersion ultrasonic emitter. During treatment, the ozonized water was injected both into the internal and external surfaces of endoscopes. The output ozone concentration reached 30 mg/l with dry air flow rate 0.3 l/min. Phenolphthalein test and "Delatest" solution method determined the remains of alkali and blood, respectively.
机译:在本文中,提出了一种特殊设计的设备,用于对装有臭氧的超声罐容器中的有机化合物(血液和碱)进行腹腔镜,内窥镜和其他长距离外科手术设备的预灭菌处理。第一个设备配备了一个浴缸(最大30升),十个总功率为300 W的超声波发射器以及四个基于介电势垒放电的模块化臭氧发生器。还设计了第二个配备3.5升容积浴的设备,该设备的玻璃管长1米,直径8厘米,带有一个臭氧模块和浸入式超声波发射器。在治疗过程中,将臭氧水注入内窥镜的内表面和外表面。输出臭氧浓度达到30 mg / l,干燥空气流量为0.3 l / min。酚酞试验和“ Delatest”溶液法分别测定了碱和血液的残留量。