首页> 外文OA文献 >Highly Flexible Silicone Coated Neural Array for Intracochlear Electrical Stimulation
【2h】

Highly Flexible Silicone Coated Neural Array for Intracochlear Electrical Stimulation

机译:高度柔性硅胶涂层神经阵列,用于胰膜卷筒电刺激

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

We present an effective method for tailoring the flexibility of a commercial thin-film polymer electrode array for intracochlear electrical stimulation. Using a pneumatically driven dispensing system, an average 232±64 μm (mean ± SD) thickness layer of silicone adhesive coating was applied to stiffen the underside of polyimide multisite arrays. Additional silicone was applied to the tip to protect neural tissue during insertion and along the array to improve surgical handling. Each array supported 20 platinum sites (180 μm dia., 250 μm pitch), spanning nearly 28 mm in length and 400 μm in width. We report an average intracochlear stimulating current threshold of 170±93 μA to evoke an auditory brainstem response in 7 acutely deafened felines. A total of 10 arrays were each inserted through a round window approach into the cochlea’s basal turn of eight felines with one delamination occurring upon insertion (preliminary results of the in vivo data presented at the 48th Annual Meeting American Neurotology Society, Orlando, FL, April 2013, and reported in Van Beek-King 2014). Using microcomputed tomography imaging (50 μm resolution), distances ranging from 100 to 565 μm from the cochlea’s central modiolus were measured. Our method combines the utility of readily available commercial devices with a straightforward postprocessing step on the order of 24 hours.
机译:我们提出了一种用于剪裁商业薄膜聚合物电极阵列的柔韧性的方法,用于Interacochear电刺激。使用气动驱动的分配系统,平均232±64μm(平均±SD)厚度层施加硅氧烷粘合剂涂层的厚度层,以加强聚酰亚胺多路阵列的下侧。将额外的硅胶施加到尖端上以在插入期间保护神经组织,并沿着阵列来改善手术处理。每个阵列都支持20个铂网站(180μmdia,250μm间距),跨越近28 mm的宽度,宽度为400μm。我们报告平均胰蛋白色刺激电流阈值为170±93μA,以引起7个急性耳聋的腰椎听觉脑干反应。总共10个阵列通过圆形窗口方法插入Cochlea的基础转弯,在插入时发生一次分层(在第48届年度会议美国神经外科学会,奥兰多,佛罗里达州的体内数据的初步结果2013年,并报道于2014年梵高国王)。使用微型仿形断层摄影成像(50μm分辨率),测量来自耳蜗中央Modiolus的100至565μm的距离。我们的方法将易于获得的商业设备的效用与24小时的顺序相结合了直接的后处理步骤。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号