First experience in implantation of a mechanical circulatory support device based on a disk-type pump: an acute experiment

机译:基于盘式泵的机械循环支撑装置植入机械循环支撑装置的第一经验:急性实验

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

Objective: to carry out the implantation of an artificial left ventricle of the heart based on a disk-type pump in an acute experiment on a large mammal (mini-pig). Materials and methods. To test the surgical technique of implantation and assess the biocompatibility of the apparatus for mechanical support of blood circulation based on a viscous friction pump, an acute experiment was conducted on an animal. A large mammal (mini-pig weighing 90 kg) was used as an experimental model. The implantation of the pump was performed extracorporeally according to the scheme «the apex of the left ventricle – the descending thoracic aorta». During the experiment, invasive blood pressure, central venous pressure, cardiac arrhythmias, body temperature, blood gas composition, activated coagulation time were monitored. Under the control of transesophageal echocardiography, the pump operation mode was set with parameters – speed 2400–2600, productivity 4 ± 0.5 l/min, average IAD – 70–80 mm Hg. Results. In the course of the experiment, the fundamental possibility of using the developed disk-type pump as a device for supporting blood circulation was proved. For 4 hours, the pump provided adequate hemodynamic parameters with an average productivity of 4 ± 0.5 l/min and 2500 rpm. After 4 hours of operation of the pump in the conditions of inactivated heparin (AST – 114 sec), no blood clots were found between the pump disks. Conclusion The hemodynamics feature of the disk pump allows you to develop sufficient performance parameters to ensure adequate blood circulation. The mechanism of action of the «boundary layer» minimizes the risk of blood clots in the pump cavity. However, the topographic and anatomical features of the pig’s body do not allow experiments with a long observation period.
机译:目的:开展人工植入左基于在急性实验盘式泵在大型哺乳动物(迷你猪)的心脏心室。材料和方法。为了测试植入手术技术和评估机械支撑基于粘性的摩擦泵血液循环装置的生物相容性,急性实验是在动物中进行。大型哺乳动物(微型猪体重90千克)被用作实验模型。按照该方案进行体外泵植入«左心室心尖 - 胸降主动脉»。在实验过程中,有创血压,中心静脉压,心律失常,体温,血液气体组成,活化凝血时间进行了监测。下食管超声心动图的控制中,泵的操作模式被设置以参数 - 速度2400-2600,生产率4±0.5升/分,平均IAD - 70-80毫米汞柱。结果。在实验的过程中,使用开发的盘型泵作为用于支持血液循环装置的基本可能性被证明。 4小时后,该泵提供了充分的血液动力学参数为4±0.5升/分钟和2500转的平均生产力。 (AST - 114秒)4小时灭活肝素的条件下,泵的操作后,被所述泵盘之间没有发现血块。结论盘泵的血流动力学功能,可以产生足够的性能参数,以保证充足的血液循环。所述«边界层»的作用机制最小化血液凝块在泵腔的风险。但是,猪的身体的地形图和解剖学特征不允许与长时间的观察期实验。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号