Adaptation and validation of a commercial head phantom for cranial radiosurgery dosimetry end-to-end audit

机译:用于颅放射前术的商业头部模型的适应与验证剂量术结束结束审计

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

Objectives: To adapt and validate an anthropomorphic head phantom for use in a cranial radiosurgery audit. Methods: Two bespoke inserts were produced for the phantom: one for providing the target and organ at risk for delineation, and one for performing dose measurements. The inserts were tested to assess their positional accuracy. A basic treatment plan dose verification with an ionisation chamber was performed to establish a baseline accuracy for the phantom and beam model. The phantom and inserts were then used to perform dose verification measurements of a radiosurgery plan. The dose was measured with alanine pellets, EBT-XD film and a plastic scintillation detector (PSD). Results: Both inserts showed reproducible positioning (±0.5 mm) and good positional agreement between them (±0.6 mm). The basic treatment plan measurements showed agreement to the Treatment Planning System (TPS) within 0.5%. Repeated film measurements showed consistent gamma passing rates with good agreement to the TPS. For 2%-2mm global gamma, the mean passing rate was 96.7% and the variation in passing rates did not exceed 2.1%. The alanine pellets and PSD showed good agreement with the TPS (-0.1 and 0.3% dose difference in the target) and good agreement with each other (within 1%). Conclusions: The adaptations to the phantom showed acceptable accuracies. The presence of alanine and PSD do not affect film measurements significantly, enabling simultaneous measurements by all three detectors. Advancements in knowledge: A novel method for thorough end-to-end test of radiosurgery, with capability to incorporate all steps of the clinical pathway in a time-efficient and reproducible manner, suitable for a national audit.
机译:目标:适应和验证拟人的头部幻影,用于颅放射外科审计。方法:为幻像制备两个定制插入物:一个用于为统计风险提供靶和器官的一个,用于进行剂量测量。测试插入物以评估其位置精度。进行具有电离室的基本治疗计划剂量验证,以建立幻象和光束模型的基线精度。然后使用幻影和插入物来对放射外科计划进行剂量验证测量。用丙氨酸颗粒,EBT-XD膜和塑料闪烁检测器(PSD)测量剂量。结果:两个插入件显示可重复定位(±0.5毫米)和它们之间的良好位置协议(±0.6 mm)。基本治疗计划测量显示对治疗计划系统(TPS)的一致性0.5%。重复的薄膜测量显示与TPS良好一致的一致伽玛通过速率。对于2%-2mm全球伽玛,平均通行率为96.7%,并且通过率的变化不超过2.1%。丙氨酸颗粒和PSD与TPS(靶标和0.3%的剂量差异)吻合良好,彼此良好(1%以内)。结论:对幻像的适应表明可接受的准确性。丙氨酸和PSD的存在不会显着影响薄膜测量,使所有三个探测器同时测量。知识进步:一种新的彻底端到端终端测试方法的放射牢,能够以较效率和可重复的方式纳入临床途径的所有步骤,适合国家审计。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号