Current limits for flowmeter resistance in metabolic carts can negatively affect exercise performance

机译:代谢推车中流量计抗性的电流限制可能对运动表现产生负面影响

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

Abstract Purpose To investigate whether a metabolic cart using a flowmeter in the upper range of accepted resistance to airflow (<1.5 cmH2O∙L−1∙s−1 for flows up to 14 L∙s−1, American Thoracic Society) negatively impacts exercise performance in healthy individuals. Methods 16 recreationally active males (age 25 ± 1 years, height 180 ± 6 cm, weight 73.5 ± 5.8 kg, all mean ± SD) performed two incremental tests on a bicycle ergometer on each of two visits, using a metabolic cart with a flowmeter of either low (Oxycon Pro) or high (Innocor) airflow resistance. Mouth pressures, gas exchange, blood lactate concentration [La−], perception of breathlessness, respiratory, and leg exertion were assessed throughout the tests. Results Tests performed with the Innocor were significantly shorter (15.3 ± 3.2 vs. 15.8 ± 3.3 min, p < 0.0001) and showed higher maximal flow resistance (1.3 ± 0.2 vs. 0.3 ± 0.0 cmH2O∙L−1∙s−1, p < 0.0001). At end‐exercise, peak oxygen consumption (−200 ± 220 ml.min−1, p < 0.0001), minute ventilation (−19.9 ± 10.5 L.min−1, p < 0.0001), breathing frequency (−5.4 ± 5.2 breaths.min−1, p < 0.0001), heart rate (−2.1 ± 3.6 bpm, p = 0.002) and [La−] (−0.7 ± 1.0 mmol.L−1, p < 0.0001), but not tidal volume (−0.1 ± 0.2 L, p = 0.172) were lower with the Innocor, while the perception of breathlessness was higher (+3.8 ± 5.1 points, p < 0.0001). Conclusions Airflow resistance in the upper range of current guidelines can significantly affect exercise performance and respiratory pattern in young, healthy males during incremental exercise. The present results indicate the need to revisit guidelines for devices used in ergospirometry.
机译:摘要目的为了研究代谢车是否在对气流的上限范围被接受的电阻的使用流量计(<1.5厘米水柱∙L-1∙S-1长达流向14升∙S-1,美国胸科协会)负面地影响锻炼表现在健康个体。方法16名娱乐性活性男性(年龄25±1岁,身高180±6厘米,体重73.5±5.8公斤,所有平均值±SD)上每两次访问的自行车测功计执行的两个增量的测试中,使用代谢车用流量计的任一低(Oxycon临)或高(Innocor)气流阻力。口压力,气体交换,血液乳酸浓度[镧],气喘,呼吸的感知,和腿部劳累整个试验进行了评估。结果进行试验与Innocor均显著较短(15.3±3.2对15.8±3.3分钟,P <0.0001)和显示出更高的最大流动阻力(1.3±0.2与0.3±0.0厘米水柱∙L-1∙S-1,第<0.0001)。在最终的运动,峰值耗氧量(-200±220 ml.min-1,P <0.0001),每分钟通气量(-19.9±10.5 L.min-1,P <0.0001),呼吸频率(-5.4±5.2次呼吸.min-1,p <0.0001),心脏速率(-2.1±3.6 BPM,p = 0.002)和[镧](-0.7±1.0 mmol.L-1,p <0.0001),但不潮气量( - 0.1±0.2 L,p = 0.172)均与Innocor降低,而呼吸困难的感觉是较高的(3.8±5.1分,p <0.0001)。结论在目前的指导方针范围的上限气流阻力可以增加运动时显著影响年轻,健康男性的运动性能和呼吸模式。目前的结果表明,需要重新审视在ergospirometry使用的设备的准则。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号