首页> 外文OA文献 >Prototype X-Ray Detectors Based on CVD Diamond
【2h】

Prototype X-Ray Detectors Based on CVD Diamond

机译:基于CVD金刚石的原型X射线探测器

摘要

X-ray detectors have been fabricated from a range of commercially-available chemical vapour deposition (CVD) diamond. They have been packaged and tested in a clinical environment, using clinical apparatus and following clinical procedures: 6 MV linear accelerator, solid water phantom, dosimeter.Some devices exhibited highly desirable characteristics, such as negligible dark currents (sub-pA), low priming doses (few Gy) and high specific sensitivities (up to 586 nC Gy-1 mm-3), demonstrating the potential of these devices as simple-to-use, small size, tissue-equivalent, sensitive x-ray dosimeters.The performance of such devices in clinical applications, such as beam profiling (both depth and off-axis) is currently in progress.
机译:X射线检测器是由一系列可商购的化学气相沉积(CVD)金刚石制成的。它们已使用临床设备和以下临床程序在临床环境中进行了包装和测试:6 MV线性加速器,固体水体模,剂量计。某些设备表现出非常理想的特性,例如可忽略的暗电流(sub-pA),低启动次数剂量(几Gy)和高比灵敏度(高达586 nC Gy-1 mm-3),证明了这些设备具有易于使用,体积小,组织等效,灵敏的X射线剂量计的潜力。目前,这种设备在临床上的应用,例如光束轮廓(深度和偏轴)正在开发中。