首页> 外文OA文献 >Electromagnetic inductance tomography (EMT): sensor, electronics and image reconstruction for a system with a rotatable parallel excitation.
【2h】

Electromagnetic inductance tomography (EMT): sensor, electronics and image reconstruction for a system with a rotatable parallel excitation.

机译:电磁感应断层扫描(EMT):具有可旋转并行激励的系统的传感器,电子设备和图像重建。

摘要

The main features and principle of operation of an electromagnetic inductance tomography system are described. The system can image the distribution of electrically conducting and/or magnetically permeable materials and is analogous to the other electrical tomography systems which measure capacitance or resistance. The paper presents a nonintrusive system with a rotatable parallel uniform excitation magnetic field which improves the detectability at the centre of the object space with respect to that near the periphery. The sensor construction, electronics design and image reconstruction are described. A novel electronic compensation scheme is used to condition the output signals from the sensor. This scheme subtracts the empty space background field measurements and consequently increases the overall dynamic range of the data acquisition system. An image reconstruction technique based on a correlation method is described and some images are presented.
机译:描述了电磁感应层析成像系统的主要特征和工作原理。该系统可以使导电和/或导磁材料的分布成像,并且类似于测量电容或电阻的其他电子断层扫描系统。该论文提出了一种具有可旋转的平行均匀激励磁场的非介入系统,该系统提高了相对于周围物体在中心的可探测性。描述了传感器的结构,电子设计和图像重建。一种新颖的电子补偿方案用于调节来自传感器的输出信号。该方案减去了空白空间背景场的测量值,因此增加了数据采集系统的整体动态范围。描述了一种基于相关方法的图像重建技术,并提出了一些图像。