首页> 美国政府科技报告 >Report of the Results of SIM.QM-P6: UV/Visible Spectrophotometry Wavelength Standard Interlaboratory Comparison
【24h】

Report of the Results of SIM.QM-P6: UV/Visible Spectrophotometry Wavelength Standard Interlaboratory Comparison

机译:sIm.Qm-p6结果报告:紫外/可见分光光度法波长长度标准实验室间比较

获取原文

摘要

A study was designed to compare wavelength axis measurements in ultraviolet and visible (uv/vis) spectrometry at several laboratories of the InterAmerican Metrology System (Sistema InterAmericano de Metrologia, or SIM) using holmium oxide solution Certified Reference Materials (CRMs). The goals of the study were to demonstrate comparability of wavelength measurements among the participants, to demonstrate the comparability between National Institute of Standards and Technology (NIST) and Centro Nacional de Metrologia (CENAM) versions of holmium oxide in perchloric acid solution CRMs, to acquire holmium oxide solution spectral data from a variety of spectrometers for use in a NIST study of wavelength assignment algorithms, and to provide a basis for a possible reassessment of the NIST-certified wavelength values. The goals were generally met, with widespread agreement of the measurements, evidence that sample origin is irrelevant, and support for re-assignment of the NIST-certified wavelength values for absorption band positions in SRM 2034.

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com