SECONDARY STANDARDS LABORATORY MEASUREMENT SYSTEM OPERATION PROCEDURE BE-04 AC VOLTAGE, O.OOOl TO 1000 VOLTS VARIABLE FREQUENCY

机译:二级标准实验室测量系统操作程序BE-04交流电压,零至1000伏变化频率

获取原文

摘要

1.1This procedure describes the operation of a secondary standard calibration system which is used to calibrate ac voltage measuring instruments by the Thermal Transfer method.n1.2The instrument under calibration will be referred to herein as the Test Instrument.n1.3The system instrunentation accuracy is given in Section 3. The overall accuracy achieved during instrument calibraion, however, must take into consideration the operator errorsnin manipulation and readout as well as the inherent system accuracy.