首页> 美国政府科技报告 >SECONDARY STANDARDS LABORATORY INSTRUMENT CALIBRATION PROCEDURE AE-16 VACUUM TUBE VOLTMETERS HEWLETT PACKARD 400D, 400H, AND 400L (USING THE RFL 1605)
【24h】

SECONDARY STANDARDS LABORATORY INSTRUMENT CALIBRATION PROCEDURE AE-16 VACUUM TUBE VOLTMETERS HEWLETT PACKARD 400D, 400H, AND 400L (USING THE RFL 1605)

机译:二级标准实验室仪器校准程序aE-16真空管电压计HEWLETT paCKaRD 400D,400H和400L(使用RFL 1605)

获取原文

摘要

1.1 This procedure describes the calibration of the Hewlett Packard 400D, 400H, or 400L Vacuum Tube Voltmeter, herein referred to as the Test Instrument.n1.2 Section 4 of this procedure will verify correct performance of the Test Instrument within the required tolerances. Normally, each recorded value will be compared to its tolerance and corrective action taken, if required, before continuing the test. However, it is sometimes desirable to delay the verification and corrective action until several related values have been recorded. Fornthis reason, the operator is cautioned not to verify the data until instructed to do so. Perform any adjustments in accordance with Section 6 and other corrective action in accordance with the manufacturer's manual and local laboratory practices. All data must be recorded on the Calibration Report form.n1.3 Read the entire procedure before starting the calibration.