首页> 美国政府科技报告 >SECONDARY STANDARDS LABORATORY MEASUREMENT SYSTEM OPERATION PROCEDURE HV-06 VOLTAGE STANDING WAVE RATIO 3/8' COAXIAL, TYPE N CONNECTOR (100 TO 1000 MC)nPREPARED PRINCIPALLY
【24h】

SECONDARY STANDARDS LABORATORY MEASUREMENT SYSTEM OPERATION PROCEDURE HV-06 VOLTAGE STANDING WAVE RATIO 3/8' COAXIAL, TYPE N CONNECTOR (100 TO 1000 MC)nPREPARED PRINCIPALLY

机译:二级标准实验室测量系统操作程序HV-06电压驻留波比3/8“同轴,N型连接器(100至1000 mC)npREpaRED pRINCIpaLLY

获取原文

摘要

This procedure describes the operation of a secondary standards laboratory calibration system which is used to measure the voltage standing wave ratio of 3/8 inch coaxial instruments having type N connectors, and in the frequency range from 100 Mc to 1000 Mc. Single-ended instruments such as signal generators and terminations or double-ended instru¬ments such as attenuators, directional couplers, or filters having a VSWR up to 10:1 can be calibrated with this system.nThe instrument under calibration will be referred to herein as the Test Instrument.nThe estimated system instrumentation accuracy is given in Section 3. The over-all accuracy achieved during instru¬ment calibration, however, must take into consideration the operator errors in manipulation and readout as well as the inherent system accuracy.