首页> 美国政府科技报告 >INSTRUMENT CALIBRATION PROCEDURE MU-02 OPTICAL TORQUE TESTER STURTEVANT MODEL OTT 500 AND 165
【24h】

INSTRUMENT CALIBRATION PROCEDURE MU-02 OPTICAL TORQUE TESTER STURTEVANT MODEL OTT 500 AND 165

机译:仪器校准程序mU-02光学扭矩测试仪sTURTEVaNT模型OTT 500和165

获取原文

摘要

This procedure describes the calibration of the Sturtevant optical torque tester, Model Ott 500 or 165, herein referred to as the Test Instrument.nThe Test Instrument must be in normal operating condi¬tion before starting the calibration.nRead the entire procedure before starting the cali¬bration.nThis procedure will verify correct performance of the Test Instrument within the required tolerances. Perform cor¬rective action in accordance with the instruction manual and local laboratory practices.nPerform this procedure in the order in which it is written.nCalibration report forms have been included at the end of this procedure. Section 5.14 and Appendix H of the Standards Laboratory Information Manual have been used for tolerance determination and related information. All data must be recorded on the calibration report form.nAll instruments used during this calibration must bear current calibration servicing identification tabs.