首页> 美国政府科技报告 >Automated Potentiometric Techniques for the On-Site Monitoring of Ions Important in Water Quality Control
【24h】

Automated Potentiometric Techniques for the On-Site Monitoring of Ions Important in Water Quality Control

机译:水质控制中重要离子现场监测的自动电位技术

获取原文

摘要

The characteristics of the nitrate electrode - fluoride electrode cell without liquid junction were investigated for applicability to nitrate ion analysis. In addition to enabling accurate nitrate analysis, the cell was found to be useful for examining the effects of interfering ions. An attempt was made to use the nitrate electrode in following nitrate ion uptake by algal cultures. The algae interfere to a certain extent with the electrode response. Using knowledge gained from research under this grant, a potentiometric nitrate analysis capability has been set up at the new University of Missouri Environmental Trace Substances Laboratory. (Author)