首页> 美国政府科技报告 >Evaluation of the Picker Capacitor Discharge Mobile X-Ray Machine,
【24h】

Evaluation of the Picker Capacitor Discharge Mobile X-Ray Machine,

机译:评估拾音器电容放电移动X射线机,

获取原文

摘要

Tests were performed on a Picker Model DGC-1010capacitor discharge mobile x-ray machine to determine accuracy of voltage and current-time indications,line voltage effects and current drain,exposure reproducibility,focal spot size and location,correspondence between the collimator and light localizer,inherent filtration,half-value layer,radiation output,and tube housing and collimator leakage. Spurious radiation was found to originiate in the tube under certain conditions. These conditions were investigated,and the observed radiation levels are reported. To determine beam quality differences between capacitor discharge and conventional equipment,comparisons were made of half-value layer and step-wedge and phantom radiographs. (Author)