首页> 美国政府科技报告 >Modification of a Commercial Atomic Absorption Spectrophotometer for Cold-Vapor Determination of Mercury
【24h】

Modification of a Commercial Atomic Absorption Spectrophotometer for Cold-Vapor Determination of Mercury

机译:商用原子吸收分光光度计的改进用于冷汞蒸汽测定汞

获取原文

摘要

One goal of the Bureau of Mines is to develop technology to insure a dependable supply of minerals to the United States. To support this goal, the Bureau's Albany Research Center has modified a commercial atomic absorption spectrophotometer to rapidly determine trace levels of mercury by cold-vapor atomic absorption spectrophotometry. Such diverse materials as ores, mill products, stream water, sediments, and flue dust have been analyzed. A very high dynamic range of 0.2 ng/ml to 1 microgram/ml of mercury in solutions is routinely analyzed by this instrument.