首页> 美国政府科技报告 >An Accuracy Statement for a Facility Used to Calibrate Static Pressure Transducers and Differential Pressure Transducers at High Base Pressure
【24h】

An Accuracy Statement for a Facility Used to Calibrate Static Pressure Transducers and Differential Pressure Transducers at High Base Pressure

机译:用于在高基础压力下校准静压传感器和差压传感器的设备的准确性声明

获取原文

摘要

A facility has been developed to calibrate pressure transducers that are used in the NBS Gas Mass Flow Facility. Both static and differential pressure transducers can be calibrated. An air dead weight tester is the standard for static transducers in the range from 3.8 to 4.5 MPa. An air dead weight tester is also the standard for the differential pressure transducers in the range of 2.5 kPa to 50 MPa; a cistern manometer. This, plus the uncertainties in the high pressure corrections to the cistern manometer and our measurement of the mercury temperature, contributes plus or minus 690 ppm to the uncertainty of the differential pressure transducer calibrations.