In-Depth Survey Report: Control Technology for Ethylene Oxide Sterilization in Hospitals at Bronson Methodist Hospital, Kalamazoo, Michigan

机译:深入调查报告:密歇根州卡拉马祖布朗森卫理公会医院环氧乙烷灭菌控制技术

获取原文

摘要

A visit was made to Bronson Methodist Hospital (SIC-8062), Kalamazoo, Michigan to evaluate methods used to protect workers from exposures to ethylene-oxide (75218) (EtO). Possible drain exposure was controlled by isolating all of the sterilizer unit except for the front panel, into a ventilated recess room. Charcoal tube measurements indicated the facility had achieved excellent control of the full shift exposures with all values obtained being less than 0.2 parts per million (ppm). While exposures for to the sterilizer operators were low, it would be possible to reduce them even more by removing some of the impediments to the load transfer process. Rehinging the doors of the aerator to the other side would enable a smoother transfer to be made. Levels in the recess room were 100ppm near the drain, and this room must be avoided during purging and for some time following purging. However, no warnings were posted and the authors recommend that this hazardous situation be corrected. The drain exhaust should be added to the auxiliary exhaust system. A differential pressure sensor should be installed on the auxiliary ventilation system and an emergency evacuation and response plan be developed and rehearsed. Face shields and gloves should be required to protect the maintenance worker who changes the supply canisters of EtO.

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号