首页> 美国政府科技报告 >2,4-Dinitrophenylhydrazine-Coated Silica Gel Cartridge Method for Determination of Formaldehyde in Air: Identification of an Ozone Interference
【24h】

2,4-Dinitrophenylhydrazine-Coated Silica Gel Cartridge Method for Determination of Formaldehyde in Air: Identification of an Ozone Interference

机译:2,4-二硝基苯肼涂层硅胶柱测定空气中甲醛的方法:鉴定臭氧干扰

获取原文

摘要

Two versions of the 2,4-dinitrophenylhydrazine method, a coated silica gel cartridge (solid) and acetonitrile impinger (solvent based), were used simultaneously to sample varied concentrations of ozone (0-770 ppb) and formaldehyde (20-140 ppb). Ozone was found to be a negative interference in the determination of formaldehyde by the 2,4-dinitrophenylhydrazine-coated silica gel cartridge method. At 120 ppb of ozone, formaldehyde at 40 ppb was under-reported by the cartridge method by 34% and at 300 ppb of ozone, formaldehyde measurements were 61% low. Greater losses were seen at higher ozone concentrations. Impinger sampling (2,4-DNPH in acetonitrile) showed no formaldehyde losses due to ozone.