首页> 美国政府科技报告 >TRACE-ELEMENT ANALYSIS OF 1000 ENVIRONMENTAL SAMPLES PER YEAR USING INSTRUMENTAL NEUTRON ACTIVATION ANALYSIS
【24h】

TRACE-ELEMENT ANALYSIS OF 1000 ENVIRONMENTAL SAMPLES PER YEAR USING INSTRUMENTAL NEUTRON ACTIVATION ANALYSIS

机译:利用仪器中子活化分析每年1000个环境样品的痕量元素分析

获取原文

摘要

The technology and methods developed at the Plum Brook Reactor to analyze 1000 samples per year and report data on as many as 56 elements are described. The manpower for the complete analysis of 20 to 24 sam¬ples per week required only 3 to 3.5 hours per sample. The solutions to problems encountered in sample preparation, irradiation, and counting are discussed. The automation of data reduction is described. Typical data on various sample matrices are presented.

著录项

  • 作者

    Dean W. Sheibley;

  • 作者单位
  • 年度 1974
  • 页码 1-27
  • 总页数 27
  • 原文格式 PDF
  • 正文语种 eng
  • 中图分类 工业技术;
  • 关键词

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com