首页> 美国政府科技报告 >Miniaturized Hollow-Waveguide Gas Correlation Radiometer (GCR) for Trace Gas Detection in the Martian Atmosphere.
【24h】

Miniaturized Hollow-Waveguide Gas Correlation Radiometer (GCR) for Trace Gas Detection in the Martian Atmosphere.

机译:用于火星大气中痕量气体检测的小型空心波导气体相关辐射计(GCR)。

获取原文

摘要

Gas correlation radiometry (GCR) has been shown to be a sensitive and versatile method for detecting trace gases in Earth's atmosphere. Here, we present a miniaturized and simplified version of this instrument capable of mapping multiple trace gases and identifying active regions on the Mars surface. Reduction of the size and mass of the GCR instrument has been achieved by implementing a lightweight, 1 mm inner diameter hollow-core optical fiber (hollow waveguide) for the gas correlation cell. Based on a comparison with an Earth orbiting CO2 gas correlation instrument, replacement of the 10 meter mUltipass cell with hollow waveguide of equivalent pathlength reduces the cell mass from approx 150 kg to approx 0.5 kg, and reduces the volume from 1.9 m x 1.3 m x 0.86 m to a small bundle of fiber coils approximately I meter in diameter by 0.05 m in height (mass and volume reductions of >99%). This modular instrument technique can be expanded to include measurements of additional species of interest including nitrous oxide (N2O), hydrogen sulfide (H2S), methanol (CH3OH), and sulfur dioxide (SO2), as well as carbon dioxide (CO2) for a simultaneous measure of mass balance.