Direct sampling ion trap mass spectrometer for the rapid analysis of volatile organic contaminants in groundwater samples

机译:直接取样离子阱质谱仪,用于快速分析地下水样品中的挥发性有机污染物

获取原文

摘要

Analytical results of volatile organic contaminant (VOC) concentrations in groundwater samples from A/M Area obtained using a direct sampling ion trap mass spectrometer (DSITMS) and a gas chromatograph (GC) with an electron capture detector (ECD) were comparable. Replicate water samples were collected at the Integrated Demonstration Site during four bi-weekly sampling periods and analyzed by both instruments. The calibration curves prepared for both the DSITMS and GC-ECD are essentially linear over the concentration range analyzed. The advantage of the DSITMS over the GC-ECD method is the rapid sample analysis time of the DSITMS. In order to validate the technology for use in the sample minimization efforts at SRS, duplicate groundwater samples from selected compliance wells were analyzed by DSITMS and GC-ECD. The DSITMS is appropriate to use for field screening water samples before offsite analysis. These analytical results will be compared with results from offsite labs contracted to perform contract lab program (CLP) analyses of regulatory samples when these analyses are available.