A method of the thermal neutron time decay constant measurement in small bounded media is presented. A 14 MeV pulsed neutron generator is the neutron source. The system of recording of a die-away curve of thermal neutrons consists of a (sup 3)He detector and of a multichannel time analyzer based on analyzer Canberra 35+ with multi scaler module MCS 7880 (microsecond range). Optimum parameters for the measuring system are considered. Experimental verification of a dead time of the instrumentation system is made and a count-loss correction is incorporated into the data treatment. An attention is paid to evaluate with a high accuracy the fundamental mode decay constant of the registered decaying curve. A new procedure of the determination of the decay constant by a multiple recording of the die-away curve is presented and results of test measurements are shown. (author). 11 refs, 12 figs, 4 tabs. (Atomindex citation 26:048741)