Effects of spatial resolution and spectral purity on transvenous coronary angiography images

机译:空间分辨率和光谱纯度对经静脉冠状动脉造影图像的影响

获取原文

摘要

Measurements have been made on the National Synchrotron Light Source (NSLS) Coronary Angiography X17B2 beamline under ideal and real imaging conditions to investigate the optimal imaging conditions for spatial resolution and spectral purity. The spatial resolution tests were performed using two multielement Si(Li) detectors (600 element, 0.5mm, pixel-pixel spacing; 1200 element, 0.25mm pixel-pixel spacing. Images were taken of phantoms containing iodine contrast agent over a wide range of incident beam absorption conditions. Patient images were also obtained using the same viewing projection with both detectors. Harmonics present in the imaging beam can be reduced by operating the superconducting wiggler source at reduced field strength. At regions of high absorption in the patient, the harmonics present can contribute to the detected signal. Iodine phantom images were obtained at a wiggler field strength of 3 Tesla (E(sub c)=13.3keV) and 4 Tesla (E(sub c)= I 7.8keV) for comparison. As before, patient images were obtained using the same projection at both wiggler fields. Results of the detector resolution and wiggler eld measurements will be presented for the phantoms as well as the patient scans.