首页> 美国政府科技报告 >Transmission grating spectrometer for plasma diagnostics
【24h】

Transmission grating spectrometer for plasma diagnostics

机译:用于等离子诊断的透射光栅光谱仪

获取原文

摘要

Radiation temperature is an important parameter in characterizing the properties of hot plasmas. In most cases this temperature is time varying caused by the short lived and/or time dependent nature of the plasma. Thus, a measurement of the radiation flux as a function of time is quite valuable. To this end the authors have developed a spectrometer that can acquire spectra with a time resolution of less than 1 ns and covers the spectral energy range from (approximately) 60 to 1,000 eV. The spectrometer consists of an entrance slit placed relatively near the plasma, a thin gold film transmission grating with aperture, a micro channel plate (MCP) detector with a gold cathode placed at the dispersion plane and an electron lens to focus the electrons from the MCP onto a phosphor coated fiber optic plug. The phosphor (In:CdS) has a response time of (approximately) 500 ps. This detector system, including the fast phosphor is similar to one that has been previously described. The spectrometer is in a vacuum chamber that is turbo pumped to a base pressure of (approximately) 5 x 10(sup 7) torr. The light from the phosphor is coupled to two streak cameras through 100 m long fiber optic cables. The streak cameras with their CCD readouts provide the time resolution of the spectrum. The spectrometer has a built in alignment system that uses an alignment telescope and retractable prism.