首页> 美国政府科技报告 >Improved operation of the Michelson interferometer ECE diagnostic on DIII-D
【24h】

Improved operation of the Michelson interferometer ECE diagnostic on DIII-D

机译:改进了迈克尔逊干涉仪ECE诊断在DIII-D上的操作

获取原文

摘要

The measurement of accurate temperature profiles is critical for transport analysis and equilibrium reconstruction in the DIII-D tokamak. Recent refinements in the Michelson interferometer diagnostic have produced more precise electron temperature measurements from electron cyclotron emission and made them available for a wider range of discharge conditions. Replacement of a lens-relay with a low-loss corrugated waveguide transmission system resulted in an increase in throughput of 6 dB and reduction of calibration error to around 5%. The waveguide exhibits a small polarization scrambling fraction of 0.05 at the quarter wavelength frequency and very stable transmission characteristics over time. Further reduction in error has been realized through special signal processing of the calibration and plasma interferograms.

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号