首页> 美国政府科技报告 >Evaluation of 1024 channel VUV-photo-diodes for soft x-ray diagnostic applications
【24h】

Evaluation of 1024 channel VUV-photo-diodes for soft x-ray diagnostic applications

机译:评估1024通道VUV光电二极管,用于软X射线诊断应用

获取原文

摘要

We tested the operation of 1024 channel diode arrays (Model AXUV-1024, from IRD, Inc.) in subdued room light to establish that they worked and to determine the direction and speed of the scan of the 1024 channels. Further tests were performed in vacuum in the HAP, High-Average-Power Facility. There we found that the bare or glass covered diodes detected primarily visible light as expected, but diodes filtered by aluminized parylene, produced a signal consistent with soft x-rays. It is probable that the spectral response and sensitivity, as discussed below, reproduce that previously demonstrated by 1 to 16 channel VUV-photodiodes; however, significantly more effort would be required to establish that experimentally. These detectors appear to be worth further evaluation where 25 w spatial resolution bolometers or spectrograph detectors of known sensitivity are required, and single-shot or 0.02-0.2s time response is adequate. (Presumably, faster readout would be available with custom drive circuitry.)