D0 Silicon Upgrade: Control Dewar Venturi Calibration Explanation for Toshiba

机译:D0硅升级:控制东芝的杜瓦文丘里校准说明

获取原文

摘要

This document is intended to explain the calibration data for the venturi, FE-3253H, which is installed in the control dewar. Further, this document will help explain how to use the venturi to make mass flow measurements during typical operating conditions. The purpose of the calibration data enclosed from the Colorado Engineering Experiment Station Inc. is to experimentally show that the venturi follows the flow equation which is enclosed as Eq. 7-36 on page 155, from the Applied Fluid Dynamics Handbook. The calibration data serves to show that the Subsonic Venturi, Serial Number 611980-18, produces results predicted by the compressible subsonic flow mass flow rate equation above and to experimentally determine the discharge coefficient C. Colorado Engineering Experiment Station Inc. ran tests at 15 independent differential pressures to conclude that use of this venturi will perform according to the mass flow rate equation. In order to verify the results from the Colorado Engineering Experiment Station Inc. we have provided you with a step-by-step procedure using the values they have chosen.