机译:使用NaI探测器进行层析伽马扫描

获取原文

摘要

The authors examined two approaches for using NaI detectors to perform transmission corrections used in the tomographic gamma scanner (TGS) and segmented gamma scanner (SGS) nondestructive assay methods. They found that a material-basis-set (MBS) fit using empirical logarithmic response spectra is quite accurate. Because this is a gross count technique, it gives sensitivities (for equal numbers of detectors) that are roughly ten times better than those obtained using Germanium detectors. The authors also found that simple continuum subtraction can be used in MBS fits using the energy-group-analysis technique only when the Pu transmission is greater than 10%. Both approaches for using NaI detectors require a knowledge of the Pu (or other) isotopics to obtain full accuracy.