The Advanced Photon Source (APS) linac beam position monitors can be used to monitor the position of a positron beam also containing electrons. To accomplish this task, both the signal at the bunching frequency of 2856 MHz and the signal at 2 x 2856 MHz are acquired and processed for each stripline. The positron beam position is obtained by forming a linear combination of both 2856- and 5712-MHz signals for each stripline and then performing the standard difference over sum computation. The required linear combination of the 2856- and 5712-MHz signals depends on the electrical calibration of each stripline/cable combination. In this paper, the calibration constants for both 2856-MHz and 5712-MHz signals for each stripline are determined using a pure beam of electrons. The calibration constants are obtained by measuring the 2856- and 5712-MHz stripline signals at various electron beam currents and positions. Finally, the calibration constants measured using electrons are used to determine positron beam position for the mixed beam case.