Performance of a Rapid-Scan Vacuum Michelson Interferometer at the NSLS

机译:在NsLs上使用快速扫描真空迈克尔逊干涉仪的性能

获取原文

摘要

A commercial Nicolet Magna series rapid-scan Michelson Fourier Transform Infrared(FTIR) was installed in a vacuum housing and integrated into the U4IR beamline at the National Synchrotron Light Source at Brookhaven National Laboratory. The frequency reference laser was mounted outside vacuum, but the moving mirror mechanism and the dynamic alignment system for the fixed mirror were in vacuum. The performance of the instrument was measured in the usual way by measuring the repeatability of data collected under specific conditions of aperture, resolution and mirror scanning velocity. We briefly discuss the beamline design, to put the interferometer in context, then present signal to noise data which we discuss in terms of both instrument performance and also storage ring stability. Under optimal conditions, the instrument has a reproducibility of 0.01% in 1 minute of measuring time at a resolution of 2 cmss.