首页> 美国政府科技报告 >Correction for the Absorption of Plutonium alpha Particles in Filter Paper Used for Dust Sampling
【24h】

Correction for the Absorption of Plutonium alpha Particles in Filter Paper Used for Dust Sampling

机译:用于粉尘取样的滤纸中钚α粒子吸收的校正

获取原文

摘要

This sample of air-borne dust collected on a filter paper when laboratory air is monitored for plutonium with the 1195 portable dust sampling unit may be regarded, for counting purposes, as a thick source with a non-uniform distribution of alpha-active plutonium. Experiments have been carried out to determine a correction factor to be applied to the observed count on the filter paper sample to correct for internal absorption in the paper and on the dust layer. From the results obtained it is recommended that a correction factor of 2 be used. (ERA citation 10:047223)