Calorimetric Measurement of the Absorbed Dose in Irradiation with an Electron or Photon Beam from a 60 MeV Electron Accelerator

机译:使用来自60 meV电子加速器的电子或光子束照射吸收剂量的量热测量

获取原文

摘要

A simple calorimetric dosimeter has been contrived to measure the radiation dose absorbed by di(2-ethylhexyl)phosphoric acid in irradiation with an electron or photon beam from a 60 MeV electron accelerator. The absorbed dose rate is related to the beam current measured with a beam-position monitor in the accelerator. The beam position monitor is usable for dosimetry in calibration with a calorimeter. (Atomindex citation 09:411417)