New Technique to Improve the Accuracy of Albedo Neutron Dosimeter Evaluations

机译:提高反照率中子剂量计评价准确度的新技术

获取原文

摘要

The calibration factor for albedo neutron dosimeters varies greatly depending upon the energy of the neutrons in the exposure. Calibration results obtained over an eight-year period at each Lawrence Livermore National Laboratory facility where neutron exposure may occur were reviewed. A stronger relationship than expected was found between the ratio of the readings of the 9-in. to 3-in. spheres and the percent thermal. Readings from personnel and albedo badges were reviewed. The readings were consistent with the use of a calibration factor for the albedo dosimeter which varies with changes in the ratio of the personnel and albedo dosimeter TLD readings. 2 references, 6 figures. (ERA citation 09:021415)