Solvent Hold Tank Sample Results for MCU-14-1172-1173: December 2014 Monthly Sample.

机译:mCU-14-1172-1173的溶剂保持罐样品结果:2014年12月每月样品。

获取原文

摘要

SRNL received one set of SHT samples (MCU-14-1172-1173 pulled on 12/30/2014) for analysis. The samples were combined and analyzed for composition. Analysis of the composite sample MCU-14-1172-1173 indicated low concentrations of the suppressor (TiDG), of the extractant (MaxCalix), and of the modifier (CS-7SB) in the solvent relative to their nominal values. This analysis confirms a downward trend of these components. No impurities were found in this solvent. The laboratory will continue to monitor the quality of the solvent in particular for any new impurity or degradation of the solvent components.