An x-ray fluorescence method for the determination of Ti, Cr, Cu and Ta in niobium oxide has been developed. Samples/standards in powder form are mixed with boric acid in the proportion of 1:1 (400 mg. each). Double layer pellets are prepared by pressing this mixture over a primary boric acid pellet. Philips PW-1220, a semiautomatic x-ray spectrometer with tungsten target x-ray tube for excitation and LiF (200) crystal for dispersion have been used. The determination range is from 0.005 to 0.1 percent for Ti and Cr, 0.01 to 0.1per cent for Cu and 0.05 to 1per cent for Ta. 6 tables. (Atomindex citation 18:001147)