The injection beamline runs over the last magnet before a long straight section and is then displaced downward 55.88 cm to the accelerator beamline. The displacement is magnetic and the final deflection onto the synchrotron orbit is by an electric kicker. The first component, the reverse septum magnet, bends the injection beam 25 degree downward. This is followed by the injection septum (20 degree bend upward) and the final injection kicker (5 degree bend upward). The septum magnets produce a peak field of 3.4 K gauss at a current of 28,000 amperes within a 0.1 msec long pulse. The electric kicker produces a field of 7.3 KV/cm with a pulse length of 0.0011 msec. The septum magnets are similar to each other in construction with a bending radium of 72.7 cm. The curvature is required to increase the effective aperture. Each magnet has a single-turn copper coil bonded to a stainless steel plate for reinforcement. This eliminates insulating material, which could be subject to radiation damage, at the septum. The stainless steel plate is welded to the magnet laminations. The current is confined to the septum by the insulation between the laminations, which are a standard core material. The total septum thickness with shield is 1.227 cm. Pulsing the magnet eliminates the need for water cooling. 2 refs., 4 figs. (ERA citation 13:030939)