首页> 美国政府科技报告 >Qualification of the Second ICS-3000 Ion Chromatograph for Use at the Defense Waste Processing Facility
【24h】

Qualification of the Second ICS-3000 Ion Chromatograph for Use at the Defense Waste Processing Facility

机译:第二台ICs-3000离子色谱仪用于国防废物处理设施的资格

获取原文

摘要

The Defense Waste Processing Facility (DWPF) requires the analysis of specific anions at various stages of its processing of high level waste (HLW). The anions of interest to the DWPF are fluoride, formate, chloride, nitrite, nitrate, sulfate, oxalate, and phosphate. The anion analysis is used to evaluate process chemistry including formic acid/nitric acid additions to establish optimum conditions for mercury stripping, reduction-oxidation (REDOX) chemistry for the melter, nitrite destruction, etc. The DWPF Laboratory (Lab) has recently replaced the Dionex DX-500 ion chromatography (IC) systems that had been used since 1998 by the first of two new ICS-3000 systems. The replacement effort was necessary due to the vendor of the DX-500 systems no longer supporting service contracts after 2008.